Our Products

Product Details

The Fluke 754 Documenting process calibrator is everything you needed and wished for in the 744 and more! HART communication capability is combined to deliver an integrated communicating calibrator. This rugged, reliable tool is ideal for calibrating, maintaining, and troubleshooting HART and other instrumentation.
The 754 does the work of several tools - sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device. You’ll fall in love with the new improved graphical screen, Li-on battery for longer life, USB port and new accessories to complete the package. 
 
For documentation, the 754 automates calibration procedures and captures your data. And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations.
Supports popular instrumentation management software. The 753 and 754, 743 and 744 work with the Fluke DPC/TRACK2™ software, and with popular programs from Honeywell Meridium, Emerson, Cornerstone, Yokogawa, Prime Technologies, Intergraph and others. It allows you to create procedures, instructions, and action lists to deliver fast, easy documentation.
The 754 is a power multifunction documenting calibrator that lets you download procedures, lists, and instructions created with software-or upload data for printing, archiving, and analysis. The 754 also features a powerful built in HART® interface capable of performing nearly all the day-to-day tasks you now perform with a separate communicator. 
Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters
Power transmitters during test using loop supply with simultaneous mA measurement
Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules
Create and run automated as-found/as-left procedures to satisfy quality programs or regulations. Record and document results
Holds up to a full week of downloaded procedures and calibration results.
Use many features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay etc.
Easy to use
Three year warranty
Bright white dual display. Read both sourced and measured parameters simultaneously.
Multi lingual interface
Rechargeable Li-Ion battery for 10 hour uninterrupted use. Includes gas gauge.
Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 1 ms.
Delivered with DPC/Track Sample software.
Compatible with many Asset Management software packages.
HART Capabilities 
The 754 is designed to take on nearly all the day-to-day tasks you now perform with a separate communicator. In fact, it offers the communication capabilities of the 375 HART communicator.
Requires no external box or second tool for everyday HART calibration and maintenance.
Offers fast HART communication.
Supports popular models of HART transmitters, with more devise-specific command support than any other HART field calibrator.
Works with multiple masters, burst mode, and multi-drop configurations.
Is easy to update as additional instruments are added and new HART versions are released.
Interrogate to determine device type, manufacturer, model, tag.
Reconfigure the sensor mapping of dual sensor temperature transmitters.
Read HART PV function and smart transmitter digital output while measuring analog mA output.
Read and write HART configuration functions to make field adjustments to PV range points, damping, and other top-level configuration settings.
Re-label smart transmitters by reading and writing to the HART tag field.
Measurement Accuracy
Voltage DC
 1 Year 2 Years
100.000 mV       0.02%+0.005 mV 0.03%+0.005 mV
3.00000 V 0.02%+0.00005 V 0.03%+0.00005 V
30.0000 V 0.02%+0.0005 V 0.03%+0.0005 V
300.00 V 0.05%+0.05 V 0.07%+0.05 V
Voltage AC
Range 
40 to 500 Hz     Resolution 1 Year 2 Years 
3.000 V 0.001 V 0.5%+0.002 V 1.0%+0.004 V
30.00 V 0.01 V 0.5%+0.02 V 1.0%+0.04 V
300.0 V 0.1 V 0.5%+0.2 V 1.0%+0.2 V
Current DC
  1 Year 2 Years
30.000 mA         0.01% + 5 uA 0.015%+7 uA
110.00 mA 0.01% + 20 uA 0.015%+30 uA
Resistance
  1 Year 2 Years
10.000 Ω           
 0.05% + 50 mΩ 0.07%+70 mΩ
100.00 Ω 0.05% + 50 mΩ 0.07%+70 mΩ
1.0000 kΩ 0.05% + 500 mΩ 0.07%+0.5 Ω
10.000 kΩ 0.1% + 10 Ω 0.15%+15 Ω
Frequency
  Resolution Accuracy (2 Years)
1.00 to 110.00 Hz 0.01 Hz 0.05 Hz
110.1 to 1100.0 Hz 0.1 Hz 0.5 Hz
1.101 to 11.000 kHz 0.001 kHz 0.005 kHz
11.01 to 50.00 kHz 0.01 kHz 0.05 kHz
 
Source Accuracy
Voltage DC
  1 Year 2 Years
100.000 mV          
 0.01%+0.005 mV 0.015%+0.005 mV
1.00000 V 0.01%+0.00005 V 0.015%+0.0005 V
15.0000 V 0.01%+0.0005 V 0.015%+0.0005 V
Current DC
  1 Year 2 Years 
22.000 mA (Source) 0.01%+ 0.003 mA 0.02%+ 0.003 mA
Current Sink (Simulate) 0.02% + 0.007 mA 0.04% + 0.007 mA
Resistance
  1 Year 2 Years
10.000 Ω              
 0.01% + 10 mΩ 0.015% + 15 mΩ
100.00 Ω 0.01% + 20 mΩ 0.015% + 30 mΩ
1.0000 kΩ 0.02% + 0.2 Ω 0.03% + 0.3 Ω
10.000 kΩ 0.02% + 3 Ω 0.03% + 5 Ω
Frequency
  2 Years
0.1 to 10.99 Hz 0.01 Hz
0.01 to 10.99 Hz 0.01 Hz
11.00 to 109.99 Hz 0.1 Hz
110.0 to 1099.9 Hz 0.1 Hz
1.100 to 21.999 kHz 0.002 kHz
22.000 to 50.000 kHz         
 0.005 kHz
Technical Data
Data Log Functions
Measure functions: Voltage, current, resistance, frequency, temperature, pressure
Reading rate: 1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length: 8000 readings (7980 for 30 or 60 readings/minute)
Ramp Functions
Source functions: Voltage, current, resistance, frequency, temperature
Rate: 4 steps/second
Trip detect: Continuity or voltage (continuity detection not available when sourcing current)
Loop Power Function
Voltage: Selectable, 26 V
Accuracy: 10%, 18 V minimum at 22 mA
Maximum current: 25 mA, short circuit protected
Maximum input voltage: 50 V DC
Step Functions
Source Functions Voltage, current, resistance, frequency, temperature
Manual Step Selectable step, change with arrow buttons
Autostep Fully programmable for function, start delay, stepvalue
Model Name Description
Fluke 754 Documenting Process Calibrator-HART including:
BC7240 battery charger
Li-on BP7240 battery pack
DPCTrack 2™ Sample Software
Instruction manual
NIST-traceable calibration report and data
Three sets of TP220 test probes with three sets of “extended tooth alligator clips
Two sets AC280 hook clips
C799 Soft Field Case
USB communication cable, Fluke 754HHC HART communications cable
 
Clips
AC220 AC220 SureGrip™ Alligator Clip Set
AC280 AC280 SureGrip™ Hook Clips
AC285 AC285 SureGrip™ Alligator Clip Set
 
Test Probes
TP74 TP74 Lantern Tip Test Probes
TP920 TP920 Test Probe Adapter Kit
 
Differential Pressure Modules
700P00 700P00 Differential Pressure Module (2.5 mbar)
700P01 700P01 Differential Pressure Module
700P02 700P02 Differential Pressure Module (70 mbar)
700P03 700P03 Differential Pressure Module (340 mbar)
700P04 700P04 Differential Pressure Module (1000 mbar)
 
Differential Pressure Module Wet
700P22 700P22 Differential Pressure Module Wet (70 mbar)
700P23 700P23 Differential Pressure Module Wet (340 mbar)
700P24 700P24 Differential Pressure Module Wet
 
Absolute Pressure Modules
700PA3 700PA3 Absolute Pressure Module (340 mbar)
700PA4 700PA4 Absolute Pressure Module
700PA5 700PA5 Absolute Pressure Module (2 bar)
700PA6 700PA6 Absolute Pressure Module (7 bar)
 
Batteries, Chargers and Adapters
BC7240 BC7240 Battery charger/Eliminator for 75x