Our Products

Product Details

In an RF lab, the R&S®FPL1000 is as indispensable as an oscilloscope or multimeter. It is a single measuring instrument for a variety of measurement tasks. It supports not only spectrum analysis, but also highly accurate power measurement with power sensors and analysis of analog and digitally modulated signals.

  • Frequency range 5 kHz to 7.5 GHz
  • SSB phase noise: typ. –108 dBc (1 Hz) at 10 kHz (1 GHz carrier) offset
  • DANL with preamplifier: typ. –166 dBm from 10 MHz to 2 GHz
  • Lightweight with small footprint
  • 40 MHz analysis bandwidth (option)
  • Internal signal generator
Frequency range R&S®FPL1003 5 kHz to 3 GHz
R&S®FPL1007 5 kHz to 7.5 GHz
Frequency resolution   0.01 Hz
Scaling standard linear
with R&S®FPL1-K54, RBW ≤ 1 MHz linear, logarithmic
Reference frequency, internal, nominal
Accuracy   (time since last adjustment · aging rate)
+ temperature drift + calibration accuracy
Aging per year standard 1 · 10–6
with R&S®FPL1-B4 OCXO reference frequency option 1 · 10–7
Temperature drift (0 °C to +50 °C) standard 1 · 10–6
with R&S®FPL1-B4 OCXO reference frequency option 1 · 10–7
Achievable initial calibration accuracy standard 5 · 10–7
with R&S®FPL1-B4 OCXO reference frequency option 5 · 10–8
Frequency readout
Marker resolution   0.01 Hz
Uncertainty   ±(marker frequency · reference uncertainty + 10 % · resolution bandwidth
+ ½ (span / (sweep points – 1)) + 1 Hz)
Number of sweep (trace) points default value 1001
range 101 to 100001
Number of measurement points with R&S®FPL1-K54, active EMI measurement 101 to 200001
Marker tuning frequency step size marker step size = sweep points span / (sweep points – 1)
marker step size = standard span / (default sweep points – 1)
Frequency counter resolution   1 Hz
Count accuracy   ±(frequency · reference uncertainty +
½ (last digit))
Display range for frequency axis   0 Hz, 10 Hz to max. frequency
Resolution   0.1 Hz
Maximum span deviation   0.1 %

 

  • R&S®FPL1003

    Order number 1304.0004.03

    Click on order no. to copy
    Order number 1304.0004.03 is now copied to your clipboard.
    FREQUENCY RANGE
    5 kHz to 3 GHz
    ANALYSIS BANDWIDTH
    40 MHz (FPL1-B40)
    DANL PREAMP ON, F < 3 GHZ
    typ. –165 dBm
    DANL PREAMP ON, F = 7.5 GHZ
    -
    SSB PHASE NOISE F = 1 GHZ, 10 KHZ OFFSET
    typ. –108 dBc (1 Hz)
  •