Brief description
Outstanding quality and innovation does not have to come with a high price tag. The R&S FPC spectrum analyzer provides unexpected performance at a budget-friendly price. It is engineered in Germany and designed to the same quality standards as highend instruments.
Key facts
RF performance engineered in Germany
10.1 WXGA (1366 × 768 pixel) display – largest and highest resolution in class
Tracking generator and independent CW signal generator
Built-in VSWR bridge
One port vector network analyzer with Smith chart display
General data |
||
Power supply |
||
AC supply |
input specifications |
100 V to 240 V AC, 50 Hz to 60 Hz, |
Power consumption |
R&S®FPC1000 |
14 W (nom.) |
R&S®FPC1500 |
19 W (nom.) |
|
Safety |
|
IEC 61010-1, EN 61010-1, UL 61010-1, |
Test mark |
|
VDE, GS, CSA, KC |
Manual operation |
||
Languages |
|
Chinese, English, French, German, Italian,
Hungarian, Japanese, Korean, Portuguese, Russian, Spanish |
Remote control |
||
Command set |
|
SCPI 1997.0 |
LAN interface |
|
10/100BASE-T, RJ-45 |
USB |
|
type B plug, version 2.0 |
Display |
||
Size |
|
10.1" |
Resolution |
|
1366 × 768 pixel |
Pixel errors |
|
< 2 pixel |
Audio |
||
Speaker |
|
internal |
USB interface |
|
type A plug, version 2.0 |
number of interfaces |
2 |
|
Mass memory |
||
Mass memory |
|
memory stick (not supplied), |
Data storage |
internal |
> 256 instrument settings and traces |
on memory stick, ≥ 1 Gbyte |
> 5000 instrument settings and traces |
|
Environmental conditions |
||
Temperature |
operating temperature range |
+10 °C to +40 °C |
storage temperature range |
–20 °C to +70 °C |
|
Climatic loading |
relative humidity |
+25 °/+40 °C at 85 % relative humidity in line
with EN 60068-2-30 |
Mechanical resistance |
||
Vibration |
sinusoidal |
EN 60068-2-6 |
random |
EN 60068-2-64 |
|
Shock |
|
40 g shock spectrum, |
EMC |
|
in line with European EMC Directive 2014/30/EU
including |
Dimensions (W × H × D) |
without feet |
396 mm × 178 mm × 147 mm |
including feet |
396 mm × 185 mm × 156 mm |
|
Weight |
|
3 kg (6.6 lb) |
Recommended calibration interval |
|
1 year |
Models |
Frequency range |
Phase noise |
DANL @ 1 GHz |
Instrument type |
R&S®FPC1000 |
5 kHz - 1/2/3 GHz |
< –92dBc (1Hz) |
<–158 dBm/Hz FPC–B22 |
Spectrum analyzer |
R&S®FPC1500 |
5 kHz - 1/2/3 GHz |
< –92dBc (1Hz) |
<–158 dBm/Hz FPC–B22 |
Spectrum analyzer, TG, VNA |