High-end functions in an extremely lightweight, compact package
The R&S FSL18 is an lightweight and compact spectrum analyzer for cost-conscious users who want the functionality of high-end instruments. The analyzer is ideal for a large number of applications in development, service and production. Despite its compact size, it offers a wealth of functions more typical of the high-end range, thus ensuring an excellent price/performance ratio. The R&S FSL18 features a tracking generator up to 18 GHz and can analyze signals with a bandwidth of 28 MHz.
Key facts
Frequency range from 9 kHz to 18 GHz (overrange 20 GHz)
Model with and without tracking generator
Signal analysis bandwidth of 28 MHz
Lightweight and compact for on-site installation, maintenance and service
Optional battery operation
General specifications |
||
Remote control |
||
LAN interface |
|
10/100BaseT, RJ-45 |
IEC/IEEE bus (GPIB) |
R&S FSL-B10 |
SCPI 1997.0 |
Display |
||
Resolution |
|
640 × 480 pixels |
Pixel failure rate |
|
< 2 × 10–5 |
Mass memory |
||
Mass memory |
|
flash disk (internal), USB memory stick |
Data storage |
|
> 500 instrument settings and traces |
Temperature |
||
|
operating temperature range |
+0 °C to +50 °C |
permissible temperature range |
+0 °C to +55 °C |
|
storage temperature range |
–40 °C to +70 °C |
|
Climatic loading |
|
+25 °C/+40 °C at 85 % relative humidity |
Mechanical resistance |
||
Vibration |
sinusoidal |
IEC 60068-2-6 |
random |
IEC 60068-2-64 |
|
Shock |
|
40 g shock spectrum, |
Power supply |
||
Input voltage range, AC, nominal |
|
100 V to 240 V |
AC supply frequency |
|
50 Hz to 400 Hz |
Input current, AC |
|
0.9 A to 0.3 A |
Input voltage range, DC, nominal |
R&S®FSL-B30 |
10 V to 28 V |
Input current, DC |
R&S®FSL-B30 |
8.0 A to 2.2 A |
Power consumption |
|
typ. 45 W, max. 65 W with all options |
Safety |
|
IEC 61010-1, EN 61010-1, UL 61010B-1, |
Test mark |
|
VDE, GS, CSA, CSA-NRTL |
EMC |
|
EMC Directive 2004/108/EC including: |
Dimensions (W × H × D) |
with handle |
408.8 mm × 158.1 mm × 465.3 mm |
without handle |
342.3 mm × 158.1 mm × 367.0 mm |
|
Weight |
without options |
< 7 kg (< 15.43 lb) |
with battery pack |
< 8 kg (< 17.64 lb) |
|
Recommended calibration interval |
|
1 year |
operation with external reference |
2 years |
Models |
Frequency range |
Phase noise |
DANL @ 1 GHz |
Max. analysis bandwidth |
Instrument type |
R&S®FSL18 |
9 Khz to 18 Ghz |
< –95 dBc (1 Hz) (f=500 Mhz, 10Khz
offset) |
< -130dBm/Hz Pre-amplifier off |
28 Mhz |
Spectrum analyzer |
R&S®FSL18 |
9 Khz to 18 Ghz |
< –95 dBc (1 Hz) (f=500 Mhz, 10Khz
offset) |
< -130dBm/Hz Pre-amplifier off |
28 Mhz |
Spectrum analyzer and TG |