Product overview: Fluke 52 II Dual Probe Digital Thermometer
The Fluke 52 II dual-input digital thermometer delivers fast response with laboratory accuracy (0.05% + 0.3°C). You can use the 52 II to measure two contact temperature inputs simultaneously on motors, insulation, breakers, pipes, corroded connections, liquids, and wires with industrial standard J, K, T, or E-type thermocouple temperature sensors. The user friendly-front panel and large, backlit display make it easy to view results from both inputs in multiple ways. And the sleep mode feature preserves battery life to give you a typical thousand hours of operation.
Key features
Dual-input digital thermometer for lab, QA, process calibration, food safety and HVAC
Shows any combination of T1, T2, T1-T2, plus MIN, MAX, or AVG on the large backlit dual display
Features fast response and laboratory accuracy (0.05% + 0.3°C)
Includes a relative time clock on MIN, MAX, and AVG to provide a time reference for major events
Specifications: 52 II B Data Logging
Thermometer with Dual Input |
||
General Specifications |
|
|
Temperature accuracy |
Above -100°C |
J, K, T, E, and N-type: ±[0.05% +
0.3°C]1 |
|
R and S-type: ±[0.05% + 0.4°C]1 |
|
Below -100°C |
J, K, E, and N-types: ±[0.20% +
0.3°C]1 |
|
|
T-type: ±[0.50% + 0.3°C] |
|
Temperature |
J-type |
-210°C to 1200°C |
|
K-type |
-200°C to 1372°C |
|
T-type |
-250°C to 400°C |
|
E-type |
-150°C to 1000°C |
|
N-type |
-200°C to 1300°C1 |
|
R and S-type |
0°C to 1767°C1 |
J-type |
-210°C to 1200°C |
|
K-type |
-200°C to 1372°C |
|
T-type |
-250°C to 400°C |
|
E-type |
-150°C to 1000°C |
|
N-type |
-200°C to 1300°C1 |
|
R and S-type |
0°C to 1767°C1 |
|
Environmental Specifications |
|
|
Operating temperature |
-10°C to 50°C |
|
Storage temperature |
-40°C to 60°C |
|
Humidity (without condensation) |
0% to 90%; 0°C to 35°C |
|
|
0% to 70%; 0°C to 50°C |
|
Safety Specifications |
|
|
Overvoltage category |
CSA C22.2 No. 1010.1 1992; EN 61010
Amendments 1,2 |
|
Agency approvals |
CE, CSA, TÜV (pending) |
|
Mechanical and General Specifications |
|
|
Size (L x W x D) |
173 x 86 x 38 mm |
|
Weight |
400 g |
|
Batteries |
3 AA batteries typical 1000-hour
life |
Models: Fluke 52 II Dual Probe Digital Thermometer
Fluke 52 IIFluke 52 II Dual Input Digital Thermometer
Includes:
Impact absorbing holster
Two 80PK-1 beaded probe thermocouples