DC Bias Source Testing System combines the impedance analyzer 6632S with a DC bias source.This system is capable of analyzing the saturation characteristics of inductors, coils, ferrites, and magnetic cores. 6632S supports Equivalent Circuit Analysis with seven different models to assist customers in analyzing the impact of parasitic components on impedance/frequency.The system supports magnetic saturation current and temperature rise current scanning analysis, providing research and development engineers with fast and accurate current curve variations.
Current and frequency graphic scanning analysis
Frequency response 100Hz-30MHz /10MHz/3MHz
DC Bias Current Output Max.640A/320A/120A/60A/20A
Equivalent Circuit Analysis-6632S
DCR Measurement function
Built-in the Permeability-μr
Long-term consecutive maximum power output
Direct Handler interfaces control through Impedance Analyzer
Support meter mode and list mode, sweep mode
Current Accuracy | 0.000A-1.000A 1%+5mA |
| 1.001A-5.000A 2% | |
| 5.001A-40.000A 3% | |
| DCR Test | ● |
| Constant power output | ● |
| Current switch | ● |
| (6632+6225 not supported) | |
| Graphic Scanning Analysis | Frequency graphic scanning analysis |
| Magnetic saturation curve (Isat) | |
| Rated current curve (Irms) |
Model: 6632S+ Series
Standard Accessories: DIP Test Fixture-DC Bias Current (F6210)