Product overview: Fluke 754 Documenting Process Calibrator-HART
Fluke 754 Documenting Process Calibrator with HART communication does the work of many tools
Whether your calibrating instruments, troubleshooting a problem, or running routine maintenance, the Fluke 754 with HART communication can help you get the job done faster. It does so many different tasks, so quickly and so well, its the only process calibrator you need to carry. This rugged, reliable integrated communicating calibrator is ideal for calibrating, maintaining, and troubleshooting HART and other instrumentation.
The 754 is a power multifunction documenting calibrator that you can use to download procedures, lists, and instructions created with software; or upload data for printing, archiving, and analysis. The powerful built-in HART interface is capable of performing nearly all the day-to-day tasks you now perform with a separate communicator.
In fact, the 754 does the work of several tools. It sources, simulates and measures pressure, temperature, and electrical signals with one rugged, hand-held device. For documentation, the 754 automates calibration procedures and captures your data. And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations. Plus, the new improved graphical screen, Li-Ion battery for longer life, USB port, and accessories help you work smarter and faster.
Key features
Specifications: Fluke 754 Documenting Process
Calibrator-HART
Measurement Accuracy
Technical Data |
|
|
|||
Data log functions |
Measure functions |
Voltage, current, resistance,
frequency, temperature, pressure |
|||
|
Reading rate |
1, 2, 5, 10, 20, 30, or 60
readings/minute |
|||
|
Maximum record length |
8000 readings (7980 for 30 or 60
readings/minute) |
|||
|
Ramp functions |
Source functions |
|||
|
|
Rate |
|||
|
|
Trip detect |
|||
Loop power function |
Voltage |
Selectable, 26 V |
|||
|
Accuracy |
10%, 18 V minimum at 22 mA |
|||
|
Maximum current |
25 mA, short circuit protected |
|||
|
Maximum input voltage |
50 V DC |
|||
Step functions |
Source functions |
Voltage, current, resistance,
frequency, temperature |
|||
|
Manual step |
Selectable step, change with arrow
buttons |
|||
|
Autostep |
Fully programmable for function,
start delay, stepvalue, time per step, repeat |
|||
|
Environmental Specifications |
|
|||
|
Operating temperature |
-10°C to +50°C |
|
||
|
Storage temperature |
-20°C to +60°C |
|
||
|
Dust/water resistance |
Meets IP52, IEC 529 |
|
||
|
Operating altitude |
3000 m above mean sea level (9842 ft) |
|
||
|
Safety Specifications |
|
|||
|
Agency approval |
CAN/CSA C22.2 No 1010.1-92, ASNI/ISAS82.01-1994,
UL3111, and EN610-1:1993 |
|
||
|
Mechanical and General Specifications |
|
|||
|
Size |
136 x 245 x 63 mm (5.4 x 9.6 x 2.5
in) |
|
||
|
Weight |
1.2 kg (2.7 lb) |
|
||
|
Batteries |
Internal Battery Pack Li-ion:
7.2V,4400mAh, 30 Wh |
|
||
|
Battery life |
>8 hours typical |
|
||
|
Battery replacement |
Replace without opening calibrator;
no tools required |
|
||
|
Side port connections |
Pressure module connector |
|
||
|
|
USB Connector to interface to your PC |
|
||
|
|
Digital instrument (HART) connector |
|
||
|
|
Connection for optional battery
charger/eliminator |
|
||
|
Data storage capacity |
1 week of calibration procedures
results |
|
||
|
90 day specifications |
The standard specification interval
for the 750 Seriesare 1 and 2 years. |
|
||
|
|
Typical 90 day measurement and source
accuracy can be estimated by dividing the one year % of reading or % of output specifications by 2. |
|
||
|
|
Floor specifications, expressed as % of full scale or counts or ohms remain
constant. |
|
Models: Fluke 754 Documenting Process Calibrator-HART
Includes: