The P3100 Models are available in single or dual piston formats for increased operating ranges. Units can be supplied in psi, bar, kgf/cm², and MPa. These robust instruments are highly accurate, quick and easy to use. Units feature a built-in priming pump for large volume applications, piston flotation indicators and a high quality screw press for fine pressure control.
The P3100 Series is the culmination of over 50 years of experience in the production and design of high performance pressure standards. With features designed to improve accuracy and performance, increase reliability and simplify operation these deadweight testers can be used to calibrate virtually any pressure sensing device, including transducers, transmitters, gauges or pressure switches.
Each instrument is supplied with a detachable lid which makes it neat, compact and easily portable. The weight masses are stored in a high-quality case with a self locking mechanism to protect them during transit.
Key features:
Pressure Ranges
Oil operated Ranges
to 20,000 psi (1400 bar)
Accuracy
Accuracy* ±0.015%
of reading (±0.008% optional)
*Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range.
Materials of Construction |
|
Standard weight material |
Series 3 non-magnetic austenitic
stainless steel |
Weight density |
7.8 g/cm³ |
Optional fractional weights |
Solution heat treated aluminum |
Weight density |
2.7 g/cm³ |
Piston material |
Tungsten carbide with nickel binder |
|
Density- 15.0 g/cm³ |
Cylinder material |
Hardened martensitic steel Tungsten
carbide (water above 500 psi, 35 bar) |
Thermal coefficients of expansion |
Oil piston/cylinder 16.5 ppm/°C Water (above 500 psi, 35 bar) 11 ppm/°C |
General |
|
Test port adaptors |
1/8, 1/4, 3/8 and ½ NPT or BSP |
Instrument weight |
36 lbs (16 kg) |
Instrument size |
17.5 W x 12 D x 8.5 H |
Mass set weight (typical) |
80 lbs (36 kg) |
Reservoir volume |
150 cc |
Screw press displacement |
5.5 cc |
Pump displacement |
4.7 cc per stroke |
O-ring seal materials |
Buna N as standard, Viton and EPDM
available |
Weight Increments |
|
Minimum Standard Weight Increments |
Up to 500 psi (35 bar): 1 psi (0.1
bar) |
Optional Fractional Weights |
Up to 500 psi (35 bar): 0.1 psi (0.01
bar) |
Operating Fluid |
|
Mineral oil |
Shell spindle oil 22, our reference
55-655, viscosity 40 cs@ 20°C |
Options |
|
PressCal Software |
Windows-based software program that
allows users to easily apply all necessary corrections to enhance the
deadweight tester performance. Calibration details are then stored and/or
used to automatically create a calibration certificate. PressCal is provided
as standard with all 0.008% instruments. |
Model Name |
Description |
P3111-BAR |
Hydraulic Deadweight Tester |
|
Oil Operated - Single PCU |
|
Pressure Range 1 to 35 bar |
P3111-KGCM2 |
Hydraulic Deadweight Tester |
|
Oil Operated - Single PCU |
|
Pressure Range 1 to 35 kgf/cm² |
P3111-PSI |
Hydraulic Deadweight Tester |
|
Oil Operated - Single PCU |
|
Pressure Range 10 to 500 psi |
P3111-KPA |
Hydraulic Deadweight Tester |
|
Oil Operated - Single PCU |
|
Pressure Range 100 to 3,500 kPa |
P3111-MPA |
Hydraulic Deadweight Tester |
|
Oil Operated - Single PCU |
|
Pressure Range 0.1 to 3.5 MPa |
P3112-BAR |
Hydraulic Deadweight Tester |
|
Oil Operated - Single PCU |
|
Pressure Range 4 to 140 bar |
P3112-KGCM2 |
Hydraulic Deadweight Tester |
|
Oil Operated - Single PCU |
|
Pressure Range 4 to 140 kgf/cm² |
P3112-PSI |
Hydraulic Deadweight Tester |
|
Oil Operated - Single PCU |
|
Pressure Range 40 to 2,000 psi |
P3112-MPA |
Hydraulic Deadweight Tester |
|
Oil Operated - Single PCU |
|
Pressure Range 0.4 to 14 MPa |